201
Performing Numerical Calculations
Calculation
type
Description
Integration
(Total)
Calculates the di󰀨erence between the area (V·s) enclosed
by the zero position (0 V position) and the part of the
signal waveform where its amplitude is positive and the
area (V·s) enclosed by the zero position (0 V position)
and the part of the signal waveform where its amplitude is
negative.
Calculates the integration between the cursors when
performing the calculation over a specic range (selected
using the A/B cursors).
S
: Integral value
n
: Total number of data points
di
:
i
th data point for channel
Δt
: Sampling period
di × Δt
S =
∑
i = 1
n
A cursor
B cursor
S
=
S
1
−
S
2
s
1
sss
2
s
1
ss
2
Integration
(Positive)
Calculates the area (V·s) enclosed by the zero position
(0 V position) and the area where the signal waveform’s
amplitude is positive.
When performing the calculation over a specic range
(selected using the A/B cursors), calculates the integration
between the cursors.
S
: Integral value
n
: Total number of data points
di
:
i
th data point for channel
Δt
: Sampling period
di × Δt
S =
∑
i = 1
n
, di
> 0
Only the area where the
amplitude is positive
A cursor
B cursor
S
=
S
1
s
1
s
1
Integration
(Negative)
Calculates the area (V·s) enclosed by the zero position
(0 V position) and the area where the signal waveform’s
amplitude is negative.
When performing the calculation over a specic range
(selected using the A/B cursors), calculates the integration
between the cursors.
S
: Integral value
n
: Total number of data points
di
:
i
th data point for channel
Δt
: Sampling period
di × Δt
S =
∑
i = 1
n
, di
< 0
Only the area where the
amplitude is negative
A cursor
B cursor
S
= −
S
2
sss
2
ss
2
Integration
(ABS)
Calculates the area (V·s) enclosed by the zero position (0 V
position) and the signal waveform.
When performing the calculation over a specic range
(selected using the A/B cursors), calculates the integration
between the cursors.
S
: Integral value
n
: Total number of data points
di
:
i
th data point for channel
Δt
: Sampling period
|di| × Δt
S =
∑
i = 1
n
A cursor
B cursor
S
=
S
1
+
S
2
s
1
sss
2
s
1
ss
2
Numerical and Waveform Calculations
HIOKI LR8450A964-03
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Hioki MEMORY HiLOGGER LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiLOGGER LR8450 and is the answer not in the manual?

Hioki MEMORY HiLOGGER LR8450 Specifications

General IconGeneral
BrandHioki
ModelMEMORY HiLOGGER LR8450
CategoryData Loggers
LanguageEnglish

Related product manuals