Eclipse Additional Information Page 6
2.1.1 peSPL to nHL correction values
ECochG/ABR15/ABR30/AMLR/Neuro/VEMP 0 dB 2-1-2 cycle
linear envelope
ALR/MMN 0 dB 25-50-25 ms
ISO 389-1:2000, ISO 389-2:1994, ISO 389-3:1994
ECochG/ABR15/ABR30/AMLR/Neuro/VEMP 0 dB
ALR/MMN 0 dB
ECochG/ABR15/ABR30/AMLR/Neuro/VEMP 0 dB
ALR/MMN 0 dB
ECochG/ABR15/ABR30/AMLR/Neuro/VEMP 0 dB
ALR/MMN 0 dB
Only toneburst correction values change for ALR & MMN testing.
For Click and CE-Chirps
®
LS, the same correction is applied.
Maximum stimulus intensity is limited to 100dBnHL by the air conduction transducers.
The ABR unit leaves the factory with nHL calibration, but it can easily be changed to peSPL values. Please
refer to the FAQ or contact your local Interacoustics distributor if you are interested in changing the calibra-
tion unit.