3.12.3 Protocol Settings for Bone Conduction ABR
The EP15/25 contains the default bone conduction protocols: Bone Conduction CE-Chirp LS, Bone Conduc-
tion 2 kHz TB, Bone Conduction 2 kHz NB CE-Chirp LS, and Bone Conduction Click.
One of the main test parameters of interest when performing a BC ABR is the Stimulus ear. Here the type of
transducer is indicated for masking, and it is possible to adjust the masking level from the dropdown. This
function is also available from the Temporary Protocol Setup.