9-1
9 Monitoring NIBP
9.1 Introduction
The MPM uses the oscillometric method for measuring the non-invasive blood pressure
(NIBP). This measurement can be used for adults, pediatrics and neonates.
Automatic non-invasive blood pressure monitoring uses the oscillometric method of
measurement. To understand how this method works, we’ll compare it to the auscultative
method. With auscultation, the clinician listens to the blood pressure and determines the
systolic and diastolic pressures. The mean pressure can then be calculated with reference to
these pressures as long as the arterial pressure curve is normal
Since the monitor cannot hear the blood pressure, it measures cuff pressure oscillation
amplitudes. Oscillations are caused by blood pressure pulses against the cuff. The oscillation
with the greatest amplitude is the mean pressure. This is the most accurate parameter
measured by the oscillometric method. Once the mean pressure is determined, the systolic
and diastolic pressures are calculated with reference to the mean.
Simply stated, auscultation measures systolic and diastolic pressures and the mean pressure is
calculated. The oscillometric method measures the mean pressure and determines the systolic
and diastolic pressures.
Connector for NIBP cuff
N
IBP start/stop key

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals