2-1
2 The Basics
2.1 Monitor Description
2.1.1 Intended Use
This patient monitor is intended to be used for monitoring, displaying, reviewing, storing
and transferring of multiple physiological parameters including ECG, heart rate (HR),
respiration (Resp), temperature (Temp), SpO
2
, pulse rate (PR), non-invasive blood pressure
(NIBP), invasive blood pressure (IBP), cardiac output (C.O.), carbon dioxide (CO
2
), oxygen
(O
2
), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS) and
respiration mechanics (RM) of single adult, pediatric and neonatal patients.
ST-segment monitoring and C.O. monitoring are restricted to adult patients only. The ICG is
only for use on adult patients who meet the following requirements: height: 122 to 229 cm,
weight: 30 to 159 kg. 12-lead ECG and BIS monitoring are not intended for neonatal
patients.
This monitor is to be used in healthcare facilities by clinical professionals or under their
direction. It is not intended for helicopter transport, hospital ambulance, or home use.
WARNING
z This patient monitor is intended for use only by clinical professionals or under
their guidance. It must only be used by persons who have received adequate
training in its use. Anyone unauthorized or untrained must not perform any
operation on it.
2.1.2 Contraindications
None.
2.1.3 Components
This patient monitor consists of a main unit, display, parameter module racks, SpO
2
sensor,
NIBP cuff, IBP cables, C.O. cables, CO
2
components, AG components, RM components,
BIS components, etc.