Monitoring NIBP
9-7
You can not display multiple sets of NIBP measurements in some screens such as the big
numerics screen and the 12-lead ECG analysis screen.
9.7.3 Setting the Pressure Unit
In the [NIBP Setup] menu, select [Unit] and toggle between [mmHg] and [kPa].
9.8 Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood vessel
and therefore help venous puncture.
1. In the [NIBP Setup] menu, verify that the [Cuff Press.] value is appropriate. Change it
if necessary.
2. Select [VeniPuncture].
3. Puncture vein and draw blood sample.
4. Reselect either [VeniPuncture], or the
hardkey on the monitor’s front, or the
on-screen QuickKey to deflate the cuff. The cuff deflates automatically after a set time if
you do not deflate it.
During measurement, the NIBP display shows the inflation pressure of the cuff and the
remaining time in venous puncture mode.
9.9 Resetting NIBP
If the blood pressure pump works incorrectly but the monitor does not alarm for it, you can
check the pump by resetting it. To reset the pump, select [Reset] from [NIBP Setup].
9.10 NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required once
every two years or when you doubt the measured NIBP. If the test failed, corresponding
prompt messages will be given. If no message is displayed, it means no leakage is detected.
Tools required:
îš„ An adult cuff

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals