Monitoring IBP
11-6
2. Vent the transducer to atmospheric pressure by turning on the stopcock to the air.
3. Press the
hardkey on the module, or, in the setup menu for the pressure (e.g. Art),
select [Art Zero >>]→[Zero]. During zero calibration, the [Zero] button appears
dimmed. It recovers after the zero calibration is completed.
4. After the zero calibration is completed, close the stopcock to the air and open the
stopcock to the patient.
NOTE
z Your hospital policy may recommend that the ICP transducer is zeroed less
frequently than other transducers.
11.7 Making the Pressure Calibration
A calibration ensures accurate pressure readings. Perform a calibration when you use a new
transducer, and at regular intervals according to your hospital policy. You require:
îš„ Standard sphygmomanometer
îš„ Balloon pump
îš„ Tubing
îš„ T-shape connector
WARNING
z Never perform the invasive pressure calibration while a patient is being monitored.
3-way stopcoc
k
Pressure transducer
Pressure adapter cable
Monitor

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals