Maintenance
25-2
25.2 Service Tasks
The following tasks are for our qualified service professionals only. Contact a qualified
service provider if your patient monitor needs the following services. Clean and disinfect
equipment to decontaminate it before testing or maintaining it.
Maintenance schedule Frequency
Safety checks according to
IEC60601-1
At least once every two years, after any repairs where the
power supply is replace, or if the patient monitor has been
dropped.
Performance assurance for all
measurements not listed below
At least once every two years, of if you doubt the measured
values.
Defibrillator synchronization At least once every two years, or as needed.
NIBP leakage test At least once every two years, or by your hospital policy.
NIBP accuracy test At least once every two years, or by your hospital policy.
NIBP calibration At least once every two years, or by your hospital policy.
CO
2
calibration and performance
test
At least once every two years, of if you doubt the measured
values.
AG calibration At least once every two years, of if you doubt the measured
values.
AG preventative maintenance
(pump check, internal fan check,
ambient pressure check and so forth
as described in the service manual)
At least once every two years, of if you doubt the measured
values.