MTS 793 Controller Hardware FlexTest GT Controller Connections
135
Cabling and Programming Model 407 Controllers
This section describes:
• Cabling and programming the FlexTest GT Controller to analog
program and feedback signals from a Model 407 Controller.
• Connecting FlexTest GT Controllers to interlock signals from a
Model 407 Controller.
• Setting up a Model 407 Controller to send and receive signals from
FlexTest GT Controllers.
Important When used as a programmer, FlexTest GT Controllers do not
have automated mode switching capabilities. Control mode
selection on these controllers do not effect the control mode
of the controller to which they are sending test commands.
How to Provide Program to a Model 407 Controller
1. Cable the 793 and 407 Controllers together.
Cable the analog inputs and outputs between the Model 493.10
chassis and a Model 407 Controller as follows:
Note Analog I/O connectors are located on the rear panel of the chassis.
A. Connect a Model 493.76 Analog Out BNC connector to the
program (analog) input on the 407 Controller. Connections
for single or multiple Model 407 Controllers are shown.
B. If you intend to monitor sensor feedback for data acquisition
or control feedback, connect the external controller’s
conditioner output to a Model 493.77 Filtered Analog Input
BNC connector. For conditioner output locations on various
MTS devices, see “Miscellaneous Conditioner Output
Connections” on page 150.
C. Make interlock connections as appropriate. Refer to
“Connecting Interlock Signals to 407 Controllers” on page
140.