MTS 793 Controller Hardware FlexTest GT Controller Connections
77
Model 493.40 I/O Carrier Connections
Each Model 493.40 I/O Carrier module can include up to four
mezzanine cards. Each installed mezzanine card is assigned a specific
I/O Carrier module front panel connector (J4–J7 I/O).
A hardware interface file (hwi) defines each type of module (and their
associated mezzanine cards) and maps each module location for the
system software. The hwi file and the physical locations for each type
of module and associated mezzanine cards must match. Also, the I/O
Carrier module address setting (on the module) must match the hwi
file address. For more information on the hwi file, see the MTS 793
Utility Software manual.
• Transducer connections require a conditioner mezzanine card to
be installed in the I/O Carrier module. The following conditioners
can be installed:
– Model 493.21B Digital Universal Conditioner
– Model 493.25 Digital Universal Conditioner
– Model 493.47 Encoder
– Model 493.48 Acceleration Conditioner
Interlock—Station per system 493.74 J43 100-007-948
Remote Station Control—with E-Stop 055-634-2XX 493.71 J50 055-634-101
Remote Station Control—with Station Stop 055-634-3XX 493.71 J50 055-634-101
Workstation Link (Ethernet cable) 051-996-8XX
10/100
‡
N/A
Remote Setpoint Adjust Control 100-051-058 493.40 J4–J7 N/A
* -XX specifies cable length. –01 through –09 represent 10–50 ft. in 5 ft. increments. Higher numbers
represent custom cable lengths
.
† Jumper plugs are required if a cable is not installed.
‡ The connector is the 10/100 BaseT connector on the power PC module.
System Cables (part 2 of 2)
C
ABLE DESCRIPTION PART
N
UMBER
*
MODULE
C
ONNECTOR
JUMPER
P
LUG
†
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com