MTS 793 Controller Hardware FlexTest GT Controller Connections
87
Excitation sense Switches S1-6 and S1-7 select the local (on board) or remote (through a
cable) excitation sensor connection.
The local selection is used with 4-wire transducer connections.
The remote selection is used with 8-wire transducer connections.
Switch S1-7 configures the + excitation signal.
Switch S1-6 configures the – excitation signal.
Both switches need to be set for the same configuration.
Single ended
excitation
Jumper W8 connects -EX to ground for situations requiring single-
ended excitation.
Remote
S1-6 OFF
S1-7 OFF
S1-6 ON
S1-7 ON
Local
Default
Default
–EX = –EX
–EX = GND
W8W8