MTS 793 Controller Hardware FlexTest IIm Controller Connections
163
Cable specification Cable (sensor)–Two twisted shielded pairs with an overall braided
shield (MTS material 003-724-401). Cable outer shield drain wire to pin
7 (optional) and cable outer braid folded back under the EMI backshell
clamp.
Cable (valve)–18 AWG, 4-conductor shielded (Belden 9418 or
equivalent). Cable outer shield drain wire to pin 7 (optional) and to the
EMI backshell.
AC sensor connection
DC sensor connection
+FB
-FB
Inner
Guard
+EX
-EX
AC Sensor
(LVDT)
Inner
Guard
Outer
Shield
4
12
2
15
1
10
7
3
5
13
-FB
-FBR
+FB
+FBR
-EX
-EXS
+EX
+EXS
DC Sensor
Drain
Braid
4-wire connections are the
most common. An 8-wire
connection is typically
used for long cables.
The additional wires for an
8-wire connection are
shown with dashed lines.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest IIm and is the answer not in the manual?

MTS Systems FlexTest IIm Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest IIm
CategoryController
LanguageEnglish

Related product manuals