MTS 793 Controller Hardware FlexTest IIm Controller Connections
163
Cable specification Cable (sensor)–Two twisted shielded pairs with an overall braided
shield (MTS material 003-724-401). Cable outer shield drain wire to pin
7 (optional) and cable outer braid folded back under the EMI backshell
clamp.
Cable (valve)–18 AWG, 4-conductor shielded (Belden 9418 or
equivalent). Cable outer shield drain wire to pin 7 (optional) and to the
EMI backshell.
AC sensor connection
DC sensor connection
+FB
-FB
Inner
Guard
+EX
-EX
AC Sensor
(LVDT)
Inner
Guard
Outer
Shield
4
12
2
15
1
10
7
3
5
13
-FB
-FBR
+FB
+FBR
-EX
-EXS
+EX
+EXS
DC Sensor
Drain
Braid
4-wire connections are the
most common. An 8-wire
connection is typically
used for long cables.
The additional wires for an
8-wire connection are
shown with dashed lines.