MTS 793 Controller Hardware
206
D/A Connections Digital-to-analog mezzanine cards support up to six program or
readout signals to external devices.
• Each D/A signal is an analog output
within ±10 volts.
• Each readout signal is from a 16 bit
digital to analog converter.
• Each output is a ±10 V analog output.
Cable specification The cable specifications apply to both the analog inputs and analog
outputs.
The cable from the front panel of the I/O Carrier module has the
following specification:
• 15 contact type D male EMI connector.
• Backshell–EMI metallized plastic.
• Cable type–up to 6 shielded twisted pairs, each with the drain
wire connected to the signal source.
Encoder Connections
See Encoder Connections 109.
+
-
+
-
+
-
Output 1
Output 2
Output 3
Output 4
Output 5
Output 6
+
-
+
-
+
-
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
J4 - J7
From
D/A
To
External
Device
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com