EasyManuals Logo
Home>MTS Systems>Controller>FlexTest IIm

MTS Systems FlexTest IIm User Manual

Default Icon
240 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #207 background imageLoading...
Page #207 background image
MTS 793 Controller Hardware
206
D/A Connections Digital-to-analog mezzanine cards support up to six program or
readout signals to external devices.
• Each D/A signal is an analog output
within ±10 volts.
• Each readout signal is from a 16 bit
digital to analog converter.
• Each output is a ±10 V analog output.
Cable specification The cable specifications apply to both the analog inputs and analog
outputs.
The cable from the front panel of the I/O Carrier module has the
following specification:
• 15 contact type D male EMI connector.
• Backshell–EMI metallized plastic.
• Cable type–up to 6 shielded twisted pairs, each with the drain
wire connected to the signal source.
Encoder Connections
See Encoder Connections 109.
+
-
+
-
+
-
Output 1
Output 2
Output 3
Output 4
Output 5
Output 6
+
-
+
-
+
-
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
J4 - J7
From
D/A
To
External
Device
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest IIm and is the answer not in the manual?

MTS Systems FlexTest IIm Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest IIm
CategoryController
LanguageEnglish

Related product manuals