EasyManuals Logo
Home>MTS Systems>Controller>FlexTest IIm

MTS Systems FlexTest IIm User Manual

Default Icon
240 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
Installing 497/498 Electronics (FTIIm)
MTS 793 Controller Hardware Installation
61
Model 497.01 Analog Chassis
About Analog Chassis Slots
Slot pairing Chassis slot assignments are grouped in pairs. If you have a
conditioner in a slot, then you cannot have a valve driver installed in
the slot paired with it. Slots are paired as follows:
Slot assignments The analog chassis contains 18 bus-board slots for plug-in modules.
Each slot is reserved for a particular type of plug-in module.
Slot A Reserved for the Model 497.34 Multistation Interlock Module.
Slot B Reserved for the Model 497.36 Communications Module.
Slots 1–4 These slots are reserved for up to four AC conditioners.
Slots 5–8 These slots are reserved for up to four DC conditioners.
Slots 9–12 These slots are reserved for options (not used with standard systems).
Slots 13–16 These slots are reserved for up to four valve drivers.
Slots 1–2
Slots 3–4
Slots 5–6
Slots 7–8
Slots 9–10
Slots 11–12
Slots 13–14
Slots 15–16
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest IIm and is the answer not in the manual?

MTS Systems FlexTest IIm Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest IIm
CategoryController
LanguageEnglish

Related product manuals