EasyManuals Logo
Home>MTS Systems>Controller>FlexTest IIm

MTS Systems FlexTest IIm User Manual

Default Icon
240 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #233 background imageLoading...
Page #233 background image
MTS 793 Controller Hardware
Optional Station Configurations
232
RSC Remote Station Control (RSC) is not supported for 6-station
configurations.
8-Station Configuration
The 8-Station configuration can provide eight channels of control.
A typical 8-Channel/8-Station configuration requires 16 Digital
Universal Conditioners and 8 two-stage valve drivers.
HSM power limits HSM power current is limited to 1.5 A per HSM.
Interlocks For 8-station configurations the hwi file must contain the line
INTERLOCKS=8. This line must be a discrete entry, not part of any
other hwi section.
Cross-head interlocks with solenoid power are not supported for 8-
station configurations.
RSC Remote Station Control (RSC) is not supported for 8-station
configurations.
A/D analog inputs Only one set of six A/D analog inputs is supported for 8-station
configurations.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest IIm and is the answer not in the manual?

MTS Systems FlexTest IIm Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest IIm
CategoryController
LanguageEnglish

Related product manuals