MTS 793 Controller Hardware
FlexTest GT Controller Connections
138
Note Because of wiring considerations between 793 and 407 Controllers,
you must assign an HSM in the station configuration file even if an
HSM does not control power to the external controller.
3. On the 793 Controller, adjust the program signals.
Use the Station Manager application to adjust the full scale,
polarity, gain, and offset for the external controller channel
(Display menu > Station Setup).
If your station configuration includes sensor feedback, you can
also adjust the full scale, gain, polarity, and offset of the incoming
sensor signals, as well as limits.
4. On the 793 Controller, set up the program.
Define a program with the Function Generator, Basic TestWare, or
Multipurpose TestWare applications.
Note It is good practice to use tapered wave shapes to apply the
program command to the external controller slowly.
The Basic TestWare application allows you to create simple test
programs that do not require complex signal management or
mode switches for station configuration files. To create more
sophisticated test programs use the optional MultiPurpose
TestWare application.
5. On the 407 Controller, select the External program source.
On the Function Generator menu, select Waveform, then scroll
through the selections to select External.