MTS 793 Controller Hardware FlexTest IIm Controller Connections
167
ADDA II Connections
Each optional Model 493.50 ADDA II module supports up to four
mezzanine cards. Each installed mezzanine card is assigned a specific
ADDA II module front panel connector (J11–J18).
A hardware interface file (hwi) defines each type of module (and their
associated mezzanine cards) and maps each module location for the
system software. The hwi file and the physical locations for each type
of module and associated mezzanine cards must match. Also, the
ADDA II module address setting (on the module) must match the hwi
file address.
The following mezzanine cards can be installed:
Model 493.55 A/D
Model 493.56 D/A
Model 493.57 DSPAD
Model 493.59 Universal Encoder
ADDA II Module I/O
Connections
The Model 493.55 A/D, Model 493.57 DSPAD, and Model 493.56 D/A
mezzanine cards can be installed in the Model 493.50 ADD II module
assembly as needed. A connection from the front panel connector to a
rear panel BNC transition module allows easy access to the analog
channels.
Note The external analog source can be a stand alone function generator,
another controller, or a computer controlled analog output.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest IIm and is the answer not in the manual?

MTS Systems FlexTest IIm Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest IIm
CategoryController
LanguageEnglish

Related product manuals