Installing the 493.02 Chassis (FlexTest SE)
MTS 793 Controller Hardware Installation
49
VMEbus Modules
The VMEbus compatible modules should be installed in chassis slots
according to the following guidelines.
• The processor module is installed in slot 1
• The I/O carrier module is installed in slot 2.
• The System I/O module is installed in slots 4 and 5.
VMEbus Modules
S
LOT MODULE NAME FUNCTION
Slot 1 Processor Provides PIDF processing for the Controller and management
of the other plug-in modules in the standalone configuration.
Provides an interface between the Controller and an external
computer in the automated configuration.
Slot 2 Model 493. 40 I/O Carrier Supports up to four mezzanine cards that can be used for
sensor conditioning, driving a servovalve, and in the
automated configuration, providing A/D and/or D/A channels.
Slot 3 Blank in standard
configurations.
May be used for additional
Model 493.40 I/O Carrier
(factory installed option for
automated systems)
Additional carrier chassis that can house up to four
mezzanine cards for expanded sensor conditioning and A/D
and D/A channels
Note This is not a standard option. It is available only for
automated systems and must be configured at the
factory. For more information, contact MTS.
Slots 4, 5 Model 493.42 System I/O Provides box-in and box-out connections for daisy chaining
multiple Controllers, analog and digital I/O, E-Stop, and HSM
and HPU control.
5
4
3
2
1
TYPE E PC-MIP
TYPE E PC-MIP
RUN 6TS BFL
ABT RST
DEBUG
PCI MEZZAN NE CARD
Placement of
VMEbus modules
in the chassis
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com