The parameter enables a user to use the drying mode settings as in case of the
moisture analyzers MAC and MAX series. Thus, on purchasing a moisture
analyzer MA 3Y series it is not necessary to determine the drying parameters for
the products to be dried, but simply use the selected and proved drying modes.
Procedure:
• Enter submenu < Others>
• Select parameter <Compatibility with MAC/MAX> which opens a window
with drop down menu,
• Select one of available options:
o YES – parameter enabled
o NO – parameter disabled
Selecting the parameter value causes immediate returning to the menu
window.
23.11. Sensors sensitivity
The parameter is adjustable between 0 – 9, and it determines the
distance from which the IR proximity sensors should react.
As standard, the sensitivity is set within the range of 5 to 7.
Procedure:
• Enter submenu < Others>
• Select parameter <Sensor sensitivity> which opens an editing window,
• Select one of available options. The selection causes immediate returning
to the menu window.
23.12. Proximity sensors delay
Parameter for setting delay of proximity sensors, it is expressed in [ms].
By default the value is set to <0>.
When using option of motion recognition the user must remember about setting <proximity
sensors delay> parameter to value <500ms>
Procedure:
a) Go to < Other> submenu.
b) Select <Proximity sensors delay> parameterand wait for a dialog
window with an on-screen keyboard to open.
c) Enter value <500>.
d) Confirm. The program returns to menu window.
23.13. Autotest
Option permanently disabled in the moisture analyzer MA 3Y series.