− Gross – main measuring unit (adjustment unit)
− Current result – currently displayed measuring unit
− Supplementary unit – supplementary measuring unit
− Mass – main measuring unit (adjustment unit)
Non-standard printout
Non-standard printout
a printout may contain: TEXTS and VARIABLES (which are acquired from
the software on printout).
Each non-standard printout is a separate project, featuring specific name,
by which it is identified, and saved in the database of printouts.
PROCEDURE:
1. Press option <Non-standard printout>
2. Press <Add> key
3. A window is opened with the following data: Name/Code/Project
4. Set name and code of a printout
5. Press <Project> key
6. A keyboard is opened on the display for editing the printout
7. Use keyboard to design printout template, the printout may contain
texts and variables.
Caution:
User can add a printout by importing configured texts from data storage devices
connected to moisture analyzer’s USB port.
Printout’s name is ONLY a NAME, and it is not part of its content.
Means of designing a non-standard printout template is described in point 19.10
<Printouts> of this user manual
16.6. Profiles
Profiles are described in point 15. <Profile> of this user manual.
17. DRYING PROCESS – GENERAL INFORMATION
Working mode < Drying> is dedicated to determining moisture content in
a substance by its drying. On process start controlled by a user, the moisture
analyzer checks loss of mass of a substance placed in the drying chamber. The
substance sample is heated by a moisture analyzer, and on basis of mass
decrement, the software calculates moisture content in a sample, and visualizes
measurement result on the display. The drying result is continuously updated
during drying process course, and the process is completed automatically on
meeting finish mode requirements set by a user.