CAUTION: Remember to start the moisture analyzer up with no load on the drying
pan.
5.1. Temperature stabilization period of time
One must wait until the thermal stability of the balance has been reached
before carrying out a weighing process.
In case of balances that were stored in low temperature (wintertime) it takes
about 4 hours to acclimatize and heat before being plugged to mains. The
balances may not be plugged to mains earlier than it has been specified by a
manufacturer.
During the stabilization process the display readouts may vary. Correct
operation of the balance is possible within a temperature range specified in
technical parameters (see point 1 of this manual).
It is recommended to find such a place for balance in operation where the
ambient temperature changes are insignificant.
5.2. Moisture analyzer switching off
CAUTION:
REMEMBER, before unplugging the moisture analyzer from mains, the
terminal must be switched off using ON/OFF key located on
terminal’s overlay!
Procedure:
• Press < > key located on terminal’s overlay
• A window is opened with the following options:
- Shut down (switching off operation of the terminal)
- Cancel (cancel procedure)
• Press < Switch off > key
• Switch off the moisture analyzer using the main switch key located at the
back of the housing
• Unplug the moisture analyzer from mains.
6. KEYBOARD – FUNCTION KEYS