- 82 -
Caution:
The checkweighing thresholds and target mass options may function
simultaneously during drying process. In such case, the tolerance field is
realized by Min and Max thresholds, and the bar graph indicates target mass
marker.
On entering the database and selecting a product, which has the target
mass value and Min & Max thresholds assigned, the software, while
controlling the start mass, will first consider the Min and Max thresholds.
The values of Min and Max thresholds can be changes before drying process
using quick access keys displayed in the bottom bar.
New values are accepted are instantaneous, and they are not saved in the
product record in a corresponding database.
18.2.3. Using database of drying programs
Drying procedure can be carried out with use of a program saved in the
<Database of drying programs>.
Adding a program to the Database of Drying Programs
Database of drying programs contains saved drying procedures, which are
easily recalled before drying a sample. Adding procedure functions easily and
very intuitively, and a user has to follow commands and hints that appear on
moisture analyzer’s display.
Procedure:
Enter submenu < Databases>, and press < Database of drying
programs>
Press < Add > key to add a new program to the database