EasyManuals Logo
Home>RADWAG>Measuring Instruments>MA 3Y

RADWAG MA 3Y User Manual

RADWAG MA 3Y
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
- 33 -
11.3. Leveling
A moisture analyzer MA 3Y series features an electronic level control system
AutoLEVEL System:
The AutoLEVEL System in MA 3Y series continuously monitors level status
during moisture analyzer operation. The level status is signaled in the upper
right corner of terminal’s display. The system continuously controls level, and
on detecting level change it indicates appropriate information on the display
and/or signals an alarm. Then the software opens a window for adjusting the
level.
Leveling procedure
Press level status pictogram < > located in the upper right corner of the
display.
The display indicates a control panel of leveling function. Moisture analyzer
operator previews level status, and instrument’s top elevation.
Level the moisture analyzer by turning the adjustable feet in a way
indicated by blinking pictograms on the display < > the level point
should move towards the center of the level circle.
As the level point is in the center point, its colour changes from red to
green which is a confirmation for correct level status.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG MA 3Y and is the answer not in the manual?

RADWAG MA 3Y Specifications

General IconGeneral
BrandRADWAG
ModelMA 3Y
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals