DS4000E User’s Guide 10-7
Chapter 10 Waveform Recording RIGOL
Tip
When the “Play back” mode is enabled, you can use SINGLE to play back the
waveform manually. Each time SINGLE is pressed, the Current Frame moves
one frame forward.
Auto Playback
After recording the waveform, press Record Mode; turn the multi-function knob
to select “Play back” and press the knob. In the waveform recording menu, you
can set the playback parameters and execute the palyback operation according to
the following instructions.
1. Specify the playback mode
Press Play Mode to select “ ” or “ ”.
: single playback. Play from the start frame to the end frame and
then stop.
: circular playback. Play from the start frame to the end frame
repeatedly until you stop it manually.
2. Set the interval
The interval refers to the time interval between the frames during the playback
process.
Press Interval and turn the multi-function knob or the inner knob of the
navigation knob to adjust the interval at a relatively smaller step or turn the
outer knob of the navigation knob to adjust the interval at a relatively larger step.
The adjustable range of the interval is from 100 ns to 10 s. The default is 100 ns.
Note: The larger the rotation amplitude of the outer knob of the navigation
knob is, the faster the variation of the value will be. When you need to adjust the
interval within a relatively larger range, this function can help you quickly finish
the setting.
3. Set the start frame
Press Start Frame and turn the multi-function knob to set the start frame
of playback. The maximum range is from 1 to the number of frames recorded.
The default is 1.
Note: The actual range of the start frame is from 1 to the end frame currently
set (End Frame).