DS4000E User’s Guide 10-15
Chapter 10 Waveform Recording RIGOL
inner knob of the navigation knob to adjust the data frame currently
displayed at a relatively smaller step or turn the outer knob of the
navigation knob to adjust the data frame currently displayed at a
relatively larger step. The adjustable range is from 1 to the number of
frame recorded. The default is 1.
Note: The larger the rotation amplitude of the outer knob of the
navigation knob is, the faster the variation of the value will be. When
you need to adjust the data frame currently displayed within a
relatively larger range, this function can help you quickly finish the
setting.
To Set the Template (Trace Mode)
In the waveform recording menu (Mode is set to Analyze), press Analyze to select
“Trace”. Then, set the template used in analysis based on trace through the method
below.
1. Specify the trace mode
Press Trace to select the creation method of analysis template.
Current Frame: select the current frame of data as the analysis template.
Average: select the average of the current data frame as the analysis
template.
2. Set the specified waveform frame as the template
Press Template to set the waveform frame currently displayed on the screen as
the waveform analysis template.
If the waveform frame currently displayed on the screen is not the one required,
you can press Current Frame and turn the multi-function knob or the
inner knob of the navigation knob to adjust the data frame currently displayed at
a relatively smaller step or turn the outer knob of the navigation knob to adjust
the data frame currently displayed at a relatively larger step. The adjustable
range is from 1 to the number of frame recorded. The default is 1.
Note: The larger the rotation amplitude of the outer knob of the navigation
knob is, the faster the variation of the value will be. When you need to adjust the
data frame currently displayed within a relatively larger range, this function can
help you quickly finish the setting.

Table of Contents