Chapter 13 System Function Setting RIGOL
DS4000E User’s Guide 13-11
Self-calibration
The self-calibration program can make the oscilloscope quickly reach the optimum
working state to acquire measurement values as accurate as possible. You can
configure the instrument to perform the self-calibration program at any time; it is
especially recommended when the temperature change is up to or more than 5.
Make sure that the oscilloscope has been warmed up or operated for more than 30
minutes before performing self-calibration.
The self-calibration uses the fast edge signal output from the [Trig
Out/Calibration] connector at the rear panel as the calibration signal. Please
connect this signal to the four input channels and the external trigger input channel,
and then press Utility Self-Cal to open the self-calibration menu and the
self-calibration interface.
Press Start and the oscilloscope will start to execute the self-calibration program.
Press Exit at any time; the oscilloscope will stop the self-calibration operation and
exit the self-calibration menu.
Note:
During the self-calibration process, the instrument will check the cable
connection status. If cable connection error occurs, Selfcal connection error!
will be displayed; please check the cable connection and restart the
self-calibration operation.
Most of the keys are disabled during the self-calibration process.