10 Probes and Biopsy Brackets
130 Basic User Manual
4. Rinse the surfaces of the probe cable with fresh running water to completely remove
the residual disinfectant.
5. Dry the probe cable with a lint-free soft dry cloth.
Table 10-4 Recommended Method for Medium-level Disinfection
Disinfectant Manufacturer Active
Ingredients
Level of Active
Ingredients
Contact
Type
Contact
Period
T-spray II Pharm. Inc. Quat.
Ammonia
/ Spray/
Wipe
<10
minutes
T-spray Pharm. Inc. Quat.
Ammonia
/ Spray/
Wipe
<10
minutes
■ Sterilization
If the probe cable comes into contact with any blood or body fluid, you should sterilize it.
For details, refer to the information about probe sterilization.
10.1.7 Storage and Transportation
You should store or transport the probe in accordance with Appendix A Specifications.
■ To Store the Probe
Follow the following items to store the probe.
●
Store the probe in a probe holder or specific package when not in use.
●
Ensure the probe holder is clean before storing it.
●
To avoid contaminating the package, disinfect or sterilize the probe before storage or
store it in a disposable probe package.
●
Do not expose it to direct sunlight or sudden environmental temperature changes.
●
Store the probe alone to avoid any impact on the probe head.
●
Carefully wind the probe cable.
●
Dry the probe before storing it.
■ To Transport the Probe
Perform the following steps to transport the probe.
1. Clean and disinfect the probe before transportation.
2. Place the probe fully into the carrying case and carefully wind the cable.
3. Close the cover of the carrying case.
4. Pack the carrying case with sponge and place it in a carton.