3 Preparing the System
26 Basic User Manual
Only use the battery if there is an emergency or no reliable AC power supply. It is
recommended to use the AC supply instead of the battery.
If you do not attempt to use the battery within 3 months, fully charge the battery and
store it in the temperatures of -20°C to 45°C. If the battery has been stored for more
than 3 months, it is recommended to charge the battery at least once every 3 months
to avoid liquid leakage.
Charge the battery in the temperature of 0°C to 45°C and discharge it in the
temperature of 0°C to 60°C to prolong the service life of the battery.
When the battery capacity is low and the battery cannot be charged in time, you
should save the data before system shutdown to avoid data loss.
3.2.3 Installing/Removing the Battery
■  To Install the Battery 
Install the battery into the battery compartment as shown in the following figure. If it is
installed in place, a click will be heard.
■  To Remove the Battery
Switch the battery lock to the position and take out the battery, as shown in the
following figure.