10 Probes and Biopsy Brackets
Basic User Manual 137
a
Biopsy Needle
Biopsy Enhanced
Region Line
Ultrasound Image
Press the function key corresponding to SteerAngle and rotate the function key to
adjust the angle of the biopsy enhanced region line. Angle a becomes more closely to
90°, and brightness of the biopsy needle becomes more stronger.
6. Gently remove the probe from the patient after completing the biopsy.
10.2.6 Cleaning the Biopsy Bracket
Clean the biopsy bracket after each use.
Wear sterile medical gloves and protective goggles during cleaning.
Perform the following steps.
1. Disconnect the probe from the system, and remove the biopsy bracket and biopsy
guide tube from the probe.
2. Brush and rinse the bracket with enzymatic cleaner to remove any residue.
3. Dry the bracket with a lint-free soft dry cloth.
10.2.7 Sterilizing the Biopsy Bracket
!
●
Disinfect the biopsy bracket after each use.
●
Use the liquid sterilant that meets local laws and regulations.
●
If you use the recommended sterilant, ensure the requirements for the sterilant level
and biopsy bracket soaking period are met. Otherwise, the biopsy bracket could be
damaged and your warranty may be void. If you have further use or needs for the
probe, ensure the level of the make-up sterilant and the biopsy bracket soaking period
are suitable for the intended clinical application.
●
Do not use the expired sterilant.
●
If you use other contact type to sterilize the biopsy bracket, ensure it is guaranteed by
the manufacturer or the local distributer.
●
Do not sterilize the biopsy bracket with ethylene oxide.
●
Do not allow any sterilant to be air-dried on the biopsy bracket.
Perform the following steps.