6 Routine Maintenance
132
Operator Manual
AMSCO® 7000 Series 10432589_G
WARNING
PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD
Always wear Personal Protective Equipment (PPE) appropriate for the task to be
performed.
Regularly scheduled routine maintenance, as well as preventive maintenance, are
required for safe and reliable operation of this equipment. Routine maintenance
procedures are explained in this manual. For preventive maintenance, contact STERIS
to schedule preventive maintenance or obtain the necessary MAINTENANCE
MANUAL if preventive maintenance is done by the Customer.
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-
trained service personnel. Repairs and adjustments performed by unqualified
personnel or installation of unauthorized parts could cause personal injury, result in
improper equipment performance, void the warranty or result in costly damage.
Contact STERIS regarding service options.
1. Regular service and maintenance MUST be performed only by STERIS or a STERIS-trained technician. Any work
performed by inexperienced or unqualified persons or the installation of unauthorized parts could cause
personal injury, invalidate the warranty or result in costly damage.
2. Under no circumstances should this equipment be serviced without the Maintenance Manual. The Maintenance
Manual can be purchased by contacting STERIS Customer Service.
3. A detailed Interval Based Checklist (Preventive Maintenance Guide) and replacement parts list can be found in
the Maintenance Manual. The Maintenance Manual can be purchased by contacting STERIS Customer Service.
4. Preventive Maintenance is essential in keeping this equipment in optimal working condition. STERIS
recommends established an annual maintenance agreement with STERIS service.
6.2 Routine Maintenance
Routine maintenance procedures included in Table 6-1 are to be performed by the Operator at indicated frequencies.
For detailed preventive maintenance procedures to be performed by a STERIS-trained service technician, refer to
Preventive Maintenance Checklist included in Maintenance Manual.
Table 6-1. Routine Maintenance - Check List
Usage and utility conditions may require more or less frequent inspections. Subsystem
1.0
EACH INSPECTION
1.1
Discuss equipment operation with department personnel.
Preparation
1.2
If printer accessory is present: Inspect printouts for signs of trouble. Preparation
1.3
Verify proper date and time are displayed; if not, reset time and date (refer to Section
5.11, Date and Time Menus).
Control
1.4
Verify buzzer is functioning properly while washer/disinfector is idling:
Press Emergency Stop pushbutton;
An alarm is generated;
Control

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris AMSCO 7053HP and is the answer not in the manual?

Steris AMSCO 7053HP Specifications

General IconGeneral
BrandSteris
ModelAMSCO 7053HP
CategoryWasher
LanguageEnglish

Related product manuals