7 Troubleshooting
186
Operator Manual
AMSCO® 7000 Series 10432589_G
7.57 ALARM# 7 SUMP TOO LONG TO HEAT (NON-CRITICAL)
Condition
Alarm is triggered when phase temperature setpoint is not reached within the allowed time (adjustable parameter).
General Troubleshooting
Additional Information
Building steam supply line shutoff valve is closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. If washer/disinfector is steam-heated: verify if
building steam supply line shutoff valve is open.
3. Press Ack on touch screen to acknowledge alarm,
then press Resume on touch screen to resume
cycle.
4. If situation reoccurs, contact STERIS.
Air supply line shutoff valve is closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. If washer/disinfector is steam-heated: verify if air
supply line shutoff valve is open.
3. Press Ack on touch screen to acknowledge alarm,
then press Resume on touch screen to resume
cycle.
4. If situation reoccurs, contact STERIS.
Building steam supply line and condensate return line
pressure do not meet requirements
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. If washer/disinfector is steam-heated: verify if
building steam supply line and condensate return
line pressure meet equipment drawing
requirements.
3. Press Ack on touch screen to acknowledge alarm,
then press Resume on touch screen to resume
cycle.
4. If situation reoccurs, contact STERIS.