7 — Troubleshooting
190
Operator Manual
AMSCO® 7000 Series 10432589_G
7.65 ALARM# 15 RECIRCULATION TEMPERATURE PROBE #2 DEFECTIVE (CRITICAL)
Condition
Alarm is triggered when during a cycle, signal from recirculation temperature probe #2 RTD (TE1 [PV]) failed.
General Troubleshooting
Additional Information
Cycle must be aborted.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Contact STERIS.
7.66 ALARM# 16 CHEMICAL INJECTED VOLUME TOO LOW (CRITICAL)
Condition
Alarm is triggered when injection phase is completed and chemical volume monitored by the flowmeter is too low.
Review alarm printout to see which chemical line triggered the alarm.
General Troubleshooting Additional Information
Chemical container is empty.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Verify if chemical container is empty. If so:
a. Replace empty container with new one.
b. Prime chemical feed line to completely
remove air that entered the system.
c. Repeat cycle.
5. If situation reoccurs, contact STERIS.