7 — Troubleshooting
181
10432589_G AMSCO® 7000 Series
Operator Manual
7.44 UNLOAD CONVEYOR NOT READY
Condition
Message is displayed when cycle has been completed and unload conveyor is still full.
General Troubleshooting
Additional Information
Conveyor needs to be emptied.
1. Empty unload conveyor.
2. Press Ack on touch screen to acknowledge
message.
7.45 UNLOAD DOOR DOES NOT OPEN
Condition
Unload door does not open.
General Troubleshooting
Additional Information
Door interlock is activated.
1. Unload door will not open until washer/disinfector
has run a complete cycle.
7.46 °F DIFFERENT ON PRINTOUT AND CAT; OR THERMAL DISINF. PASSED AS SHOWN
IN PROCONNECT
Condition
Value in Fahrenheit is not the same on printout and on CAT (ConnectAssure Technology); Or When a cycle contains only
a Drying phase, Thermal Disinfection “PASS” is reported to be consistent with ProConnect data.
General Troubleshooting
Additional Information
Temp. values different on washer/disinfector data versus
connectivity data.
1. Cycle graphs, cycle printout and thermal rinse
records can be used to confirm temperature
reached during cycle. These files can be
downloaded from the washer/disinfector.
2. Rely on washer/disinfector data first, not on
connectivity data.