6 — Routine Maintenance
138
Operator Manual
AMSCO® 7000 Series 10432589_G
6.7 Processing DECONTAMINATION Cycle
WARNING
CHEMICAL BURN AND/OR EYE INJURY HAZARD
● Chemicals are caustic and can cause adverse effects to exposed tissues. Do not get in
eyes, on skin or attempt to swallow. Read and follow the precautions and
instructions on the chemical label and in the Safety Data Sheet (SDS) prior to
handling the chemical, refilling the chemical container or servicing the chemical
injection pump. Wear appropriate Personal Protective Equipment (PPE) whenever
handling the chemical or servicing the chemical injection pump and lines.
● Wear gloves and eye protection when using a chemical. Avoid contact with the eyes
or skin. If a chemical spills or splashes on you, flush the affected area with water for
15 minutes. If swallowed, DO NOT induce vomiting. Seek medical attention
immediately.
IMPORTANT: As a preventive maintenance measure, it is the responsibility of the Customer to perform the decontamination of
the washer/disinfector once a week, as prescribed.
NOTE: Contact STERIS for information on specific cleaning and descaling products recommended for use with this
washer/disinfector.
Decontaminate and descale chamber and Acu Wash Reservoir (if present) as follows, using DECONTAMINATION cycle
expressly designed for this purpose:
1. Wear appropriate Personal Protective Equipment (PPE) such as waterproof gloves and eye protection.
2. Press Open Door on touch screen to open load door.