7 — Troubleshooting
182
Operator Manual
AMSCO® 7000 Series 10432589_G
7.47 ENTIRE PARAMETERS OF LONG CYCLE NOT PRINTED OR REBOOT AFTER DOUBLE/LONG
PRINT JOBS
Condition
Washer does not print entire parameters of a long cycle; Or system error; Or control reboot following double or long
printer jobs.
General Troubleshooting
Additional Information
Entire parameters of a long cycle not printed; system
error; control reboot.
1. User can print each cycle separately. If a single
cycle is too long to be printed from Supervisor
mode, cycle execution printout in its extended
version can be used to record cycle configuration.
7.48 WASHER NOT CALIBRATED
Condition
Message is displayed if a sensor, a pump or a flowmeter is not calibrated.
General Troubleshooting Additional Information
Acknowledge message and schedule a visit from
technician.
1. Press Ack on touch screen to acknowledge
message and return to Ready Screen.
2. Contact STERIS to schedule a visit of a service
technician to perform calibration.
7.49 WASHER STUCK IN PHASE: PREPARING FOR STANDBY, DRAINING RESERVOIR...
Condition
Washer stuck in phase: "Preparing for Standby, draining reservoir..."
General Troubleshooting Additional Information
User needs to allow time to complete or cancel.
1. User can either allow time for Standby to complete
its draining sequence or can cancel the Standby
request. The next time a cycle is ran, a specific
alarm will be displayed if the fault condition is still
present.