7 — Troubleshooting
165
10432589_G AMSCO® 7000 Series
Operator Manual
This section also describes the reminders and warnings which may appear on the washer/disinfector touch screens and
indicates probable causes.
This section describes the types of washer/disinfection malfunctions which may occur and indicates probable causes.
If unable to correct the problem with the use of the charts, or if a problem occurs that is not described on the charts,
please call STERIS. Service charges may be incurred. Consult your warranty for details.
NEVER ALLOW UNQUALIFIED PERSONS TO SERVICE THE AMSCO ®7052HP SINGLE-CHAMBER
WASHER/DISINFECTOR AND/OR THE AMSCO® 7053HP SINGLE-CHAMBER WASHER/DISINFECTOR.
IMPORTANT: When any alarm condition occurs, washer/disinfector and conveyor (if present) operations automatically stop.
7.3 ABNORMAL CONDUCTIVITY READINGS FOR THERMAL RINSE DRAINING
Condition
Abnormal conductivity readings for thermal rinse draining.
General Troubleshooting
Additional Information
Water supply conductivity is incorrect.
1. Contact building maintenance to test water
supplies.
Cycle program is incorrect.
1. Cycle program does not include sufficient rinsing
for the load type. Customize cycle to include an
additional rinse phase before thermal rinse.
Conductivity meter is miscalibrated.
1. Contact STERIS.
Lubricant (if applicable) or detergent pumps
miscalibrated.
1. Contact STERIS.
Draining time is to short.
1. Contact STERIS.