7 — Troubleshooting
193
10432589_G AMSCO® 7000 Series
Operator Manual
7.72 ALARM# 22 EMERGENCY STOP PRESSED (CRITICAL)
Condition
Alarm is triggered when Emergency Stop pushbutton is pressed.
General Troubleshooting
Additional Information
Emergency Stop pushbutton was pressed.
1. Pull out Emergency Stop pushbutton.
2. Verify both E-stops are pulled out.
3. Press Silence Buzzer on touch screen to stop alarm
buzzer.
4. Press Ack on touch screen to acknowledge alarm.
5. Press Abort on touch screen to abort cycle.
7.73 ALARM# 23 PRESSURE TRANSMITTER DEFECTIVE (CRITICAL)
Condition
Alarm is triggered when during a cycle, signal from pressure transmitter is failed.
General Troubleshooting Additional Information
Cycle must be aborted.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Contact STERIS.