7 — Troubleshooting
194
Operator Manual
AMSCO® 7000 Series 10432589_G
7.74 ALARM# 24 LOAD DOOR DID NOT OPEN (NON-CRITICAL)
Condition
Alarm is triggered when load door is detected as closed after automatic door opening sequence.
General Troubleshooting
Additional Information
Shutoff valve is closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify if air supply line shutoff valve is open.
4. If situation reoccurs, contact STERIS.
Door gutter is obstructed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify there is no obstruction in door gutter.
4. If situation reoccurs, contact STERIS.
7.75 ALARM# 25 LOAD DOOR DID NOT CLOSE (NON-CRITICAL)
Condition
Alarm is triggered when load door is not detected as closed after automatic door closing sequence (twelve seconds).
General Troubleshooting Additional Information
Shutoff valve is closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify if air supply line shutoff valve is open.
4. If situation reoccurs, contact STERIS.
Door path is obstructed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify there is no object obstructing door path and
try to close door again.
4. If situation reoccurs, contact STERIS.