7 — Troubleshooting
185
10432589_G AMSCO® 7000 Series
Operator Manual
7.55 ALARM# 5 SUMP NOT DRAINED (NON-CRITICAL)
Condition
Alarm is triggered when sump water level is still reached after draining allowed time is elapsed.
General Troubleshooting
Additional Information
Air supply shutoff valve closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Verify if air supply line shutoff valve is open.
3. Press Ack on touch screen to acknowledge alarm,
then press Resume on touch screen to resume
cycle.
4. If situation reoccurs, contact STERIS.
7.56 ALARM# 6 WATER REMAINED IN CHAMBER (CRITICAL)
Condition
Alarm is triggered when sump filling takes less than ten seconds and there is water remaining in chamber after draining
of previous stage.
General Troubleshooting Additional Information
Air supply shutoff valve closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Verify if air supply line shutoff valve is open.
5. Repeat cycle.
6. If situation reoccurs, contact STERIS.