7 — Troubleshooting
180
Operator Manual
AMSCO® 7000 Series 10432589_G
7.41 SUMP SAFETY LEVEL TEST DUE
Condition
Reminder is displayed when number of cycles set for sump safety level test (979) has been reached.
General Troubleshooting Additional Information
User needs to perform sump safety level test.
1. Press Ack on touch screen to acknowledge
reminder.
2. If you want to perform test, answer YES and
control automatically goes into sump safety level
test menu (refer to Section 6.14, Sump Safety Level
Test).
3. If you do not want to perform test, answer NO and
control automatically returns to Ready screen.
4. Reminder is displayed again after 25 more cycles
completed.
7.42 TEST FAILED: CALL STERIS
Condition
Message is displayed when automatic water inlet dynamic pressure test was unable to complete in the allowed time.
General Troubleshooting
Additional Information
Verify pressure adjustment and repeat test.
1. Verify tested water inlet pressure adjustment with
building maintenance.
2. Repeat test and if situation reoccurs, contact
STERIS.
7.43 TOUCH SCREEN DOES NOT WORK PROPERLY
Condition
Touch screen does not work properly.
General Troubleshooting
Additional Information
User needs to restart washer/disinfector.
1. Restart washer/disinfector using Power (ON/OFF)
toggle switch.
Faulty touch screen electrical connection.
1. Contact STERIS.
Touch screen lost calibration.
1. Contact STERIS.
Touch screen damaged.
1. Contact STERIS.