7 Troubleshooting
191
10432589_G AMSCO® 7000 Series
Operator Manual
7.67 ALARM# 17 RECIRCULATION TEMP. PROBES VARIANCE EXCEEDS RANGE (CRITICAL)
Condition
Alarm is triggered when recirculation temperature probe (RTD) readings are showing a variance over +2.0°C, -2.0°C for
five seconds.
General Troubleshooting Additional Information
Cycle must be aborted.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Repeat cycle.
5. If situation reoccurs, contact STERIS.
7.68 ALARM# 18 MOTOR OVERLOAD DETECTED (CRITICAL)
Condition
Alarm is triggered when performing a cycle and motor overload is detected (drying fan or recirculation pump).
General Troubleshooting Additional Information
Cycle must be aborted.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Contact STERIS.
7.69 ALARM# 19 POWER FAILURE DURING CYCLE (ABORT PROGRESS...)
Condition
Alarm is triggered when power supply was lost while performing a cycle.
General Troubleshooting Additional Information
System needs to reboot.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Wait for system to reboot.
4. If situation reoccurs, contact STERIS.