7 Troubleshooting
208
Operator Manual
AMSCO® 7000 Series 10432589_G
7.101 ALARM# 53 RINSE RESERVOIR TEMPERATURE PROBE DEFECTIVE (CRITICAL)
Condition
Alarm is triggered when signal from reservoir temperature probe (RTD) failed.
General Troubleshooting
Additional Information
Cycle must be aborted.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Press Abort on touch screen to abort cycle.
4. Contact STERIS.