6 — Routine Maintenance
156
Operator Manual
AMSCO® 7000 Series 10432589_G
Dot Identification
Position of Spray Arm Assembly
•••
•••••
Top Rotary Spray Arm Assembly
•••••••
••••••••
Bottom Rotary Spray Arm Assembly
6.13 Door Safety Test
IMPORTANT: Supervisor password is required to perform this procedure.
Every 750 cycles, a door safety test is required to verify both doors security. While washer/disinfector is in Ready
mode, an audible signal sounds and a message is displayed to remind the supervisor that the safety test must be
performed. If the supervisor decides to bypass the safety test, the message pops up again after 20 cycles. The
message will be displayed after every 20 cycles until the safety test is performed.
1. Display shows door safety test due reminder.
.
NOTE: Pressing No will bypass door safety test.
2. Press Yes to perform test.
3. Display requests supervisor password entry.
4. Enter Supervisor mode password using available touch pads on keyboard screen.
5. Press OK to confirm entry.
6. Display shows message requesting to open load door. Press Open Door to open load door. Load door opens.
7. Display shows message requesting activation of first door safety switch.
8. Insert an insulated screwdriver (shaft length of at least 51 mm [2"] and shaft diameter of 3 mm [1/10"] or less)
into right hole of detection bar.
NOTE: Detection bar is located on bottom of door. Two small holes are located under detection bar, in the middle
of the bar.