7 Troubleshooting
171
10432589_G AMSCO® 7000 Series
Operator Manual
7.17 EXCESSIVE FOAM IN CHAMBER
Condition
Excessive foam in chamber.
General Troubleshooting Additional Information
Wrong type of detergent, too foamy.
1. Refer to detergent specifications and follow
detergent manufacturer’s recommendations for
quantity of detergent to be used.
Too much detergent is injected during treatment.
1. Call supervisor to adjust detergent injection rate.
Chemical injection pumps are not calibrated properly.
1. Chemical pump needs calibration. Contact STERIS.
7.18 HAND-HELD BAR CODE READER (EMEA OPTION) DOES NOT SCAN
Condition
Hand-Held bar code reader (EMEA option) does not scan.
General Troubleshooting
Additional Information
User needs to restart washer/disinfector.
1. Restart washer/disinfector using Power (ON/OFF)
toggle switch.
2. If situation reoccurs, contact STERIS.
7.19 BAR CODE READER DOES NOT OPERATE PROPERLY
Condition
If washer/disinfector is equipped with a SCS conveyor system, bar code reader does not operate properly (cycle must
be manually selected even if a cycle bar code tag is present).
General Troubleshooting
Additional Information
Bar code tag not properly located.
1. Cycle bar code tag is not properly located on rack.
2. If situation reoccurs, contact STERIS.
Bar code reader is not located properly or is defective.
1. Bar code reader is not located properly or is
defective. Contact STERIS.