4 Operating Instructions
60
Operator Manual
AMSCO® 7000 Series 10432589_G
4.4 Loading Washer/Disinfector With a Washer Automation System
WARNING
PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD
Always load empty baskets on appropriate loading cart or surface.
Do not place empty baskets on the conveyor modules to load soiled articles. If you
do, the proximity sensor will detect the baskets and indexing will automatically start.
Maximum operating weight, including weight of basket and load, is 61 kg (135 lb) for
the AMSCO 7052HP Single-Chamber Washer/Disinfector and 73 kg (160 lb) for the
AMSCO 7053HP Single-Chamber Washer/Disinfector.
If an obstruction is present in the wash chamber door, the door safety bar will detect
an obstruction and the door will automatically stop from closing. Wait until door is
fully open before removing obstruction.
This procedure is used if the washer/disinfector is equipped with an ATLAS Workflow Automation Vehicle or with a
SCS Conveyor System and the automation is set on load side of washer/disinfector.
If automation is set only on unload side of washer/disinfector, refer to Section 4.3, Loading Washer/Disinfector Manually.
Ensure appropriate cycle parameters are set in the control before beginning operation of the washer/disinfector.
IMPORTANT: For optional ATLAS Workflow Automation Vehicle, refer to the ATLAS WAV System Operator Manual for more
detailed information about the System and transportation procedure.
IMPORTANT: For optional SCS Conveyor System, refer to SCS Conveyor System Operator Manual for more detailed information
about non-motorized or motorized loading and transporting procedure.
1. Place items to be processed in basket or accessory rack.
2. Select desired cycle bar code tag depending on load to be processed (see Table 4-1).
IMPORTANT: Always use GENTLE cycle with microsurgery instruments.
Table 4-1. Cycle Names and Cycle Bar Code Tag Numbers
CODE CYCLE NAME USE
00
INSTRUMENT
To process surgical instruments
01
UTENSIL
To process bedpans, trays, basins and bowls
04
PLASTIC GOOD
To process plasticware
05
ANESTHESIA
To process anesthesia tubing, bags and masks
06
GENTLE
To process delicate surgical instruments
07
RIGID MIS
To process rigid minimally invasive surgical instruments
08
ORTHOPEDIC
To process orthopedic instruments