312
Measurement range, maximum resolution, measurable range, and measurement accuracy gures
Measurement
target
Type Range
Maximum
resolution
Measurable range
Measurement
accuracy
Voltage – 10 mV f.s. 500 nV ±10 µV
20 mV f.s. ±20 µV
100 mV f.s. ±50 µV
200 mV f.s. ±100 µV
1 V f.s. ±500 µV
2 V f.s. ±1 mV
10 V f.s. ±5 mV
20 V f.s. 1 mV ±10 mV
100 V f.s. 5 mV ±50 mV
1-5 V f.s. 1 V to 5 V ±5 mV
Thermocouple
(Not including
accuracy of
reference
junction
compensation)
100°C f.s. 0.01°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 100°C
±0.5°C
500°C f.s. 0.05°C
±1.4°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 500°C
±0.5°C
2000°C f.s. 0.1°C
±1.4°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than 500°C
±0.5°C
Greater than or equal to 500°C
and less than or equal to 1350°C
±0.7°C
J 100°C f.s. 0.01°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 100°C
±0.5°C
500°C f.s. 0.05°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 500°C
±0.5°C
2000°C f.s. 0.1°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 1200°C
±0.5°C