313
Measurement
target
Type Range
Maximum
resolution
Measurable range
Measurement
accuracy
Thermocouple
(Not including
accuracy of
reference
junction
compensation)
E 100°C f.s. 0.01°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 100°C
±0.5°C
500°C f.s. 0.05°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 500°C
±0.5°C
2000°C f.s. 0.1°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 1000°C
±0.5°C
T 100°C f.s. 0.01°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 100°C
±0.5°C
500°C f.s. 0.05°C
±1.4°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 400°C
±0.5°C
2000°C f.s. 0.1°C
±1.4°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 400°C
±0.5°C
N 100°C f.s. 0.01°C
and less than 0°C
±1.1°C
Greater than or equal to 0°C and
less than or equal to 100°C
500°C f.s. 0.05°C
±2.1°C
and less than 0°C
±1.1°C
Greater than or equal to 0°C and
less than or equal to 500°C
2000°C f.s. 0.1°C
±2.1°C
and less than 0°C
±1.1°C
Greater than or equal to 0°C and
less than or equal to 1300°C
www.GlobalTestSupply.com