314
Measurement
target
Type Range
Maximum
resolution
Measurable range
Measurement
accuracy
Thermocouple
(Not including
accuracy of
reference
junction
compensation)
R 100°C f.s. 0.01°C 0°C to 100°C ±4.4°C
500°C f.s. 0.05°C Greater than or equal to 0°C and
less than 100°C
±4.4°C
Greater than or equal to 100°C
and less than 300°C
Greater than or equal to 300°C
and less than or equal to 500°C
±2.2°C
2000°C f.s. 0.1°C Greater than or equal to 0°C and
less than 100°C
±4.4°C
Greater than or equal to 100°C
and less than 300°C
Greater than or equal to 300°C
and less than or equal to 1700°C
±2.2°C
S 100°C f.s. 0.01°C 0°C to 100°C ±4.4°C
500°C f.s. 0.05°C Greater than or equal to 0°C and
less than 100°C
±4.4°C
Greater than or equal to 100°C
and less than 300°C
Greater than or equal to 300°C
and less than or equal to 500°C
±2.2°C
2000°C f.s. 0.1°C Greater than or equal to 0°C and
less than 100°C
±4.4°C
Greater than or equal to 100°C
and less than 300°C
Greater than or equal to 300°C
and less than or equal to 1700°C
±2.2°C
B 2000°C f.s. 0.1°C Greater than or equal to 400°C
and less than 600°C
±5.4°C
Greater than or equal to 600°C
and less than 1000°C
±3.7°C
Greater than or equal to 1000°C
and less than or equal to 1800°C
±2.4°C
C 100°C f.s. 0.01°C 0°C to 100°C ±1.7°C
500°C f.s. 0.05°C 0°C to 500°C ±1.7°C
2000°C f.s. 0.1°C 0°C to 2000°C ±1.7°C
www.GlobalTestSupply.com