Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-121
digio.trigger[N].release()
This function releases an indefinite length or latched trigger.
Usage
digio.trigger[N].release()
Digital I/O trigger line (1 to 14)
Details
Releases a trigger that was asserted with an indefinite pulsewidth time. It also releases a trigger that
was latched in response to receiving a synchronous mode trigger. Only the specified trigger line is
affected.
Example
digio.trigger[4].release()
Releases digital I/O trigger line 4.
Also see
digio.trigger[N].assert() (on page 11-116)
digio.trigger[N].pulsewidth (on page 11-120)
digio.trigger[N].reset()
This function resets trigger values to their factory defaults.
Usage
digio.trigger[N].reset()
Digital I/O trigger line (1 to 14)
Details
This function resets the following attributes to factory default settings:
â–ª digio.trigger[N].mode
â–ª digio.trigger[N].pulsewidth
â–ª digio.trigger[N].stimulus
It also clears digio.trigger[N].overrun.