Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-297
Example
node[3].setglobal("x", 5)
Sets the global variable x on node 3 to the value of 5.
Also see
node[N].getglobal() (on page 11-296)
TSP advanced features (on page 10-58)
opc()
This function sets the operation complete status bit when all overlapped commands are completed.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
opc()
Details
This function causes the operation complete bit in the Standard Event Status Register to be set when
all previously started local overlapped commands are complete.
Note that each node independently sets its operation complete bits in its own status model. Any
nodes that are not actively performing overlapped commands set their bits immediately. All remaining
nodes set their own bits as they complete their own overlapped commands.
Example
opc()
waitcomplete()
print("1")
Output:
1
Also see
Status model (on page 9-33, on page C-1)
waitcomplete() (on page 11-450)
os.time()
This function generates a time value in UTC time.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
utcTime = os.time()
utcTime = os.time(timespec)
utcTime
Time value in UTC time
timespec
The date and time (year, month, day, hour, and minute)

Table of Contents

Other manuals for Keithley 3700A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700A Series and is the answer not in the manual?

Keithley 3700A Series Specifications

General IconGeneral
BrandKeithley
Model3700A Series
CategoryMultimeter
LanguageEnglish

Related product manuals